| 	 | 	Easidew PRO XP for Gases	 | Easidew PRO XP LQ for Liquids	 | 
 	| 	Performance Specifications	 | 		
 	| 	Measurement Range	 | 	-110 to +20°C dew point -100 to +20°C dew point | 0 to 300 ppmW 
Non-standard available upon request | 
 	| 	Accuracy	 | 			±1°C dew point (+20 to -60°C) ±2°C dew point (-60 to -110°C) | 
 	| 	Response Time	 | 	5 mins to T95 (dry to wet) | 
 	| 	Repeatability	 | 	0.5°C dew point | 
 	| 	Calibration	 | 		Traceable 13 point calibration and certificate | 
 	| 	Electrical Specifications	 | 	
 	| 	Output Signal	 | 		4-20 mA (2-wire connection, current source); User configurable over range | 
 	| 	Output	 | 			Dew point, moisture content in ppmV |  Moisture content in ppmW | 
 	| 	Analog Output Scaled Range	 | 		Dew point: -110 to +20°C Moisture content in gas: 0-3000 ppmV Non-standard available upon request | 		Moisture content in liquid: 0-300 ppmW Non-standard available upon request | 
 	| 	Supply Voltage | 		14 to 28 V DC | 	
 	| 	Load Resistance | 		Max 250 Ω @ 14 V (500 Ω @ 24 V) | 	
 	| 	Current Consumption | 		23 mA max, depending on output signal | 	
 		Saturation Constants (for moisture in liquids measurements only) | 	 | 6-point look-up table for saturation constants up
to 300 ppmW over the temperature range 0 to +50°C; Saturation constants for 8 common liquids can be programmed into the Easidew PRO XP LQ via the application software; Alternatively the user can program saturation constants manually | 	
 	| 	CE Conformity | 		2004/108/EC, 94/9/E ATEX directive | 	
 	| 	Operating Specifications	 | 	
 	| 	Operating Temperature	 | 			-20 to +70°C | 
 	| 	Operating Pressure	 | 	45 MPa (450 barg) maximum | 
 	| 	Flow Rate	 | 			1 to 5 Nl/min mounted in standard sampling block; 0 to 10 m/sec direct insertion | 
 	| 	Temperature Coefficient	 | 			Temperature compensated across operating temperature range | 
 	| 	Mechanical Specifications	 | 		
 	| 	Ingress Protection | 			IP66 in accordance with standard BS EN 60529:1992; NEMA 4 protection in accordance with standard NEMA 250-2003 | 
 	| 	Explosion and Flameproof Area
Certificates	 | 			ATEX: II 2 GD Exd [ia] IIC T6 Gb EX tb IIIC T80°C Db IP66 Tamb -20°C to +70°C
  IECEx: Exd [ia] IIC T6 Gb Ex tb IIIC T80°C Db IP66 Tamb -20°C to +70°C
  CCSAUS: CLS I, Div1, GRPS ABCD CLS II & III, Div1, GRPS EFG CLS I, Zone 1 AEx/Exd [ia] IIC T6 Zone 21, AEx/Ex tb IIIC T80°C Tamb = -20°C to +70°C IP66
  TC TR EX-Certificate: Pending			 | 
 	| 	Russian Pattern Approval	 | 			[GOST-R, pending]			 | 
 	| 	Canadian Pressure Vessel Certification | 			C.R.N. - all Canadian provinces			 | 
 	| 	Housing material	 | 		Standard: Aluminium powder epoxy coated / polyurethane RAL 5009 Optional: 316 stainless steel (Housing supplied with EN10204 3.1 material cert. if
option F2 requested) | 
 		Filter (sensor protection)	 | 		Standard: Stainless steel sintered guard (for protection against fine particulate >80μm) Optional: HDPE guard (for protection against fine particulate >10μm) | 
 	| 	Process Connection and
Material	 | 			3/4” - 16 UNF with recessed Viton® o-ring 316 stainless steel			 | 
 	| 	Weight	 | 		Aluminium: 1.6kg 316 Stainless Steel: 2.4kg	 | 
 	| Electrical Connections | 	Dual 3/4” NPT gland | 
 	Diagnostic Conditions (factory programmed) | 			| Condition | Output |  
					| Sensor fault | 23 mA |  
					| Under-range dew point | 4 mA |  
					| Over-range dew point | 20 mA |  
					
  |