XTP601 Series

Oxygen Analyzer for Safe or Hazardous Areas
A robust, linear and stable oxygen analyzer for measurements in gases such as hydrogen, nitrogen, biogas, or carbon dioxide. The sensor is housed in an IP66 rugged casing, making it suitable for a wide range of applications. When supplied with flame arrestors, the analyzer becomes explosion-proof and certified for use in hazardous areas.



Product Features

  • General purpose or ATEX, IECEx, CCSAUS, TC, TR & Ex hazardous area certified models
  • Ranges available: 0–0.5% up to 0–50% & 20, 80 or 90–100%
  • Touch-screen display allows calibration or adjustment without the need for a hot works permit
  • Intrinsic error (accuracy) is better than 1% full scale
  • Zero span stability better than 0.25% of span per month
  • Outputs include RS485 and two 4-20mA (following NAMUR NE 44 standard)
  • Low cost of ownership due to minimal maintenance and no consumable parts
XTP601

Technical Specification

Performance
Measurement TechnologyThermo-paramagnetic oxygen sensor
GasProcess and non-condensing sample with particles <3µm
Measurement RangeSelectable from 0–0.5% up to 0-50% and 20, 80 or 90–100%
Display Resolution0.01% for suppressed zero ranges
Display TypeBacklit LCD
Accuracy Excluding Suppressed Zero Ranges<±1% of span or ±0.02% O2 whichever is greater
Accuracy For Suppressed Zero Ranges (80/90 - 100%)< 0.2% O2
Response Time (T90) with High Speed Response (HSR) Enabled< 15 seconds
Repeatability±0.2% of span or 0.01% O2 whichever is greater
Linearity±0.5% of span or 0.05% O2 whichever is greater
Zero Stability±0.25% of span per month
Span Stability±0.25% of span per month
Sample Flow Rate100 to 600 ml/min
Sample Flow EffectGP1 and GP3: 100 to 500 ml/min (0.2 to 1.06 scfh)
EX1, EX3 and GP2: 270 to 330 ml/min (0.2 to 1.06 scfh)
Sample Pressure0.75 to 5 BarA
Sample Temperature5 to +55°C max (+45°C for standard cell)
Sample Cell TemperatureStandard +50°C (Optional +55°C and +60°C) for ranges 0-0.5% up to 0-50% only
Background GasAnalyzer is calibrated in the background gas of the process. Standard gases are: N2, CO2, CH4, H2 & Biogas. Others available on request
Electrical Specifications
Analog Inputs2 off 4-20 mA inputs
One for an external sensor that can be displayed on the screen
One to act as an active compensation for the process conditions
Analog Outputs2 off 4-20 mA outputs (powered with 24V excitation voltage)
Output RangesPrimary range is set to the calibrated range of the instrument
The second is user selectable 0-100%
Alarms2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max)
DataloggingThe user can use the digital communications to log the output from the analyzer
The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp
Digital CommunicationsMODBUS RTU over RS485 Protocol
Power Supply24 V DC; 1.5 A max
Cable entriesAnalyzer is supplied with 3 x M20 cable entries. Cable glands, conduit entries and blanking plugs are available as accessories
Operating Conditions
Ambient Temperature5 to +55°C
Atmopheric pressure750mbar to 1250mbar
Mechanical Specifications
Warm Up Time<25 minutes
Stabilization Time5 minutes
Dimensions234 x 234 x 172mm (w x d x h)
Weight9.7kg
Wetted Materials316 & 430F stainless steel, boroscillicate glass, platinum, 3M 2216, plus O-ring
O-Ring MaterialsViton, Silicone or Ekraz
Gas ConnectionEX1, EX3 & GP2: 1/8” NPT (female)
GP1 and GP3: 1/4” NPT (female)
Ingress ProtectionIP66, NEMA 4X
Hazardous Area Classification
ATEXII 2GD Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
IECExEx d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
CCSAUSClass I, Division 1, Groups B,C,D, T6
INMETROEx d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
TC TR1 Ex d IIB+H2 T6 Gb
Temperature Ranges for ATEX and IECEx as per O-ring TypeSilicon: Ta = -40°C to +55°C
Viton®: Ta = -15°C to +55°C
Ekraz: Ta = -10°C to +55°C
The XTP601 process oxygen analyzer meets or exceeds all relevant clauses in BS EN 50104 : 2010
“Electrical apparatus for the detection and measurement of oxygen”.

Related Products

XGA301 XGA301
Bench mount analyzer for measuring oxygen, CO2, CH4, CO, O3 and others as well as dew point.
XZR500 XZR500
The XZR500 oxygen analyzer from Michell Instruments is designed to determine the excess air required for optimum combustion. It uses MSRS Zirconium Oxide technology to measure levels of oxygen in harsh conditions such as boilers, incinerators and furnaces, providing fast, accurate readings in high temperatures and corrosive atmospheres. The Pitot-effect sample probe negates the need for aspirator air in most applications.
XZR400 Series XZR400 Series
The XZR400 oxygen analyzer measures trace oxygen in pure gases. It is used to control the purity of gases such as Nitrogen, Argon, Helium and Carbon Dioxide during manufacture, quality in air separation plants, or for quality assessment of bought gases. It is available in wall-mountable and rack-mountable versions, both suited to a range of applications.
XTC601 Binary Gas Analyzer XTC601 Binary Gas Analyzer
A robust, linear and stable thermal conductivity analyzer for measurement of binary gas mixes such as air in hydrogen, nitrogen, argon, helium or carbon-dioxide.
PCR PCR
The PCR from Michell is a transportable gas blender in a flight case, suitable for calibrating a multitude of gas analyzers at virtually any location and has been designed to allow the blending of two or three gases to the required concentration.
XZR200 XZR200 Oxygen Analyzer
A cost effective zirconium-dioxide analyzer to measure percentage level oxygen in combustion processes, ambient air monitoring and many more applications.
Easidew PRO XP Easidew PRO XP