GPR-1800/2800 AIS/IS

 Intrinsically Safe ATEX Analyzers
GPR-1800 ATEX Intrinsically Safe Oxygen Analyzer from AII
The GPR-1800 measures ppm levels of oxygen and offers optional Modbus RTU Communication and modular sampling system. It is available with alarm relays (GPR- 1800 AIS) or without (GPR-1800 IS).

GPR-2800 ATEX Intrinsically Safe Oxygen Analyzer from AII
The GPR-2800 measures percent oxygen and offers optional Modbus RTU Communication and modular sampling system. As with the GPR 1800, it is available with or without alarm relays.





Produktmerkmale

  • 2 sensor options available
  • 12 to 28 V DC power or mains power (AIS only)
  • 2 x field selectable concentration alarms
  • Modbus communication (AIS only)
  • Standard sample includes sample/span valve, moisture filter, flow meters
  • Optional fully automated sample system
GPR-1000/1100/1200/2000/3500

Technische Spezifikation

Accuracy<2% of full scale range under constant conditions
Analysis Ranges GPR-1800  0-10 ppm, 0-100 ppm, 0-1000 ppm, 0.1%, 0-25% full scale ranges; auto-ranging or manually lock on a single range
GPR-2800  0-1%, 0-5%, 0-10%, 0-25% full scale ranges; auto-ranging or manually lock on a single range
Approvals Certified for use in hazardous areas UL:
United States: UL 1203, UL 913, UL 508
Canada: CAN/CSA C22.2 No. 30-M1986, CAN/CSA C22.2 No. 157-92, CAN/CSA C22.2 No. 14-10
ATEX: Directive 94/9/EC
Area ClassificationATEX certified, EX II 2 G, Exd [ib] IIB T4,
Tamb -20 to + 50°C
AlarmsTwo user configurable alarms - AIS versions only
Calibration3 month interval using air or certified span gas with O2 value approx 80% of full scale range balance N2
CompensationTemperature as standard; barometric pressure - AIS versions only
Connections1/8” compression tube fittings
ControlsWater resistant keypad; menu driven range selection, calibration and system functions
DisplayGraphical LCD 70 x 35mm; resolution 0.01 ppm; displays real time ambient temperature and pressure
EnclosureNEMA type 3R suitable for rain in outdoor applications
FlowNot flow sensitive; recommended flow rate 0.4 to 0.9 Nl/min
Linearity±1% of full scale
PressureInlet  Regulate to 0.3 to 2 barg to deliver 0.4 to 0.9 Nl/min flow;
Vent  Atmospheric
Power12 to 28 V DC
110 to 220 V AC (AIS version only)
Response Time90% of final reading in 10 seconds
Sample SystemSample flow meter; options available
Sensitivity<0.5% of full scale range
Sensor ModelGPR-1800  GPR-12-333 or XLT-12-333 (for gases containing >0.5% CO2)
GPR-2800  GPR-11-32 or XLT-11-24 (for gases containing >0.5% CO2)
Sensor LifeGPR-1800  24 months in normal operating conditions O2 @ +25°C and 1 atmosphere
GPR-2800  GPR-11-32: 32 months in normal operating conditions @ +25°C and 1 atmosphere; XLT-11-24: 24 months in normal operating conditions @ +25°C and 1 atmosphere
Signal Output4-20 mA; Modbus RTU communication (AIS only)
Temperature RangeGPR Sensor  -10 to +45°C
XLT Sensor  -20 to +45°C
Warranty12 months analyzer
12 months sensor
Wetted PartsStainless steel

Anwendungen

Automobilindustrie  Atemluft  Druckluft  Industrielle Gase  Militär / Verteidigung 
Energieerzeugung  Forschung & Entwicklung  Erdgas & Petrochemie  Pharma / Nahrungsmittel  Deponie  Chemische Industrie  Metallveredelung und Thermoprozesse

Verwandte Produkte

XZR500 XZR500
Der XZR500 Sauerstoffanalysator eignet sich für die Bestimmung des Sauerstoffüberschusses zur Regelung der optimalen Verbrennung.
XZR400TS XZR400TS
Der XZR400 Sauerstoff-Analysator eignet sich für die Messung von Sauerstoffspuren in Gasen und wird zur Überwachung der Reinheit von z.B. Stickstoff, Argon, Helium und Kohlendioxid eingesetzt.
XTP601 XTP601 Serie
Robuster, linearer und stabiler Sauerstoff-Analysator für die Messung in Stickstoff, Luft, Wasserstoff und Gasgemischen. Der Sensor sitzt in einem kompakten.