Accuracy | <2% of full scale range under constant conditions |
Analysis Ranges | 0-10*, 0-100, 0-1000, 1% and 0-25% (for calibration) |
Approvals | See individual models on Datenblatt |
Area Classification | General Purpose or ATEX approved |
Calibration | 3 month interval using certified span gas (preferred for fastest online time) or air with O2 value approximating 80% of full scale range balance N2 |
Compensation | Temperature compensation as standard Barometric pressure compensation on analyzer versions (models denoted with an ‘a’) |
Connections | 1/8” compression tube fittings |
Controls | Water resistant keypad; menu driven range selection, calibration and system functions |
Display | Graphical LCD 70 x 35mm; resolution .01 ppm |
Enclosure | See individual models on Datenblatt for material, size and weight |
Flow Sensitivity | None between 0.2 to 2.4 Nl/min 0.5 to 0.95 Nl/min recommended |
Linearity | ±1% of full scale |
Pressure | Inlet Regulate to 0.3 to 2 barg to deliver 0.5 to 0.9 Nl/min flow to transmitter Vent Atmospheric |
Power | 12 to 24 V DC, loop powered for transmitters 12 to 28 V DC or 110 to 220 V AC for analyzers |
Response Time | 90% of final full scale reading in 10 seconds |
Sensitivity | <0.5% of full scale range |
Sensor Model | GPR-1500/GPR-1500N/GPR-1500A/GPR-1500 ATEX/GPR-1500N ATEX GPR-12-333 or XLT-12-333* GPR-1500D/GPR-1500DN/GPR-1500DA/GPR-1500D ATEX/GPR-1500DN ATEXGPR-12-100-M or XLT-12-100-M *Not available with 4 range models |
Sensor Life | 24 months in <1000 ppm O2 @ +25°C and 1 atmosphere |
Signal Output | 4-20 mA |
Temperature Range | -10 to +45°C (GPR sensor) -20 to +45°C (XLT sensor) |
Warranty | 12 months analyzer 12 months sensor |
Wetted Parts | Stainless steel |
Optional Equipment | |
Sample Conditioning | Pump, filter, scrubbers - contact Michell Instruments |