| Performance | |
| Measurement range | -100 to +20°C (-148 to +68°F) dew point |
| Accuracy | ±2°C (±3.6°F) dew point |
| Response time | 5 mins to T95 (dry to wet) |
| Repeatability | 0.5°C (0.9°F) dew point |
| Calibration | Traceable 7-point calibration certificate |
| Electrical Specifications | |
| Output signal | 4-20 mA (2-wire connection, current source) User configurable over range |
| Output | Dew point, moisture content for ppmV and ppmW |
| Analog output scaled range | Dew point: -100 to +20°C (-148 to +68°F) Moisture content in gas: 0-3000 ppmV Moisture content in liquid: 0-3000 ppmW |
| Supply voltage | 12 to 28 V DC |
| Load resistance | Max 250 Ω @ 12 V (500 Ω @ 24 V) |
| Current consumption | 20 mA max depending on output signal |
| Saturation constants (For moisture in liquids measurements only) | 6-point look-up table for saturation constants up to 3000 ppmW over the temperature range 0 to +50°C. Saturation constants for 8 common liquids can be programmed into the Easidew PRO I.S. via the application software. Alternatively the user can program saturation constants manually. |
| Operating Specifications | |
| Operating humidity | 0–100% RH |
| Operating temperature | -40 to +70°C (-40 to +158°F) |
| Operating pressure | 45 MPa (450 barg / 6500 psig) max |
| Overpressure rating | x2 operating pressure 90 MPa (900 barg /
13053 psig) |
| Flow rate | 1 to 5 Nl/min (2.1 to 10.6 scfh) mounted in standard sampling block; 0 to 10 m/sec (0 to 32.8 ft/sec) direct insertion |
| Temperature coefficient | Temperature compensated across operating temperature range |
| Mechanical Specifications | |
| Ingress protection | IP66 in accordance with standard BS EN 60529:1992, and NEMA 4 in protection accordance with standard NEMA 250-2003 |
| Hazardous Area Certificates | ATEX - II 1 G Ex ia IIC T4 (-20°C ≤ Ta ≤
+70°C)
FM - IS / I / I / ABCD / T4 Ta = +70°C
CSA - IS Class I Div I Groups ABCD T4
IECEx - Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C) |
| Housing material | 316 stainless steel |
| Transmitter dimensions | L=132mm x ø27mm (L=5.19" x ø1.06") (sensor only) |
| Filter (sensor protection) | Standard: HDPE Guard <10μm
Optional: 316 stainless steel sintered guard
<80μm |
| Sensor connection and material | 5/8" - 18 UNF 316 stainless steel |
| Weight | 8 lbs |
| Interchangeability | Fully interchangeable transmitter |
| Electrical connections | Hirschmann GDS series (DIN 4350-C) |
Diagnostic conditions (factory programmed) |
| Condition | Output |
| Sensor fault | 23 mA |
| Under-range dew point | 4 mA |
| Over-range dew point | 20 mA |
|
| Approved galvanic isolators |
| KFD2-CR-EX1.20200 | KFD0-CS-EX2.50P |
| KFD2-CR-EX1.30200 | KFD2-STC4-EX1.H |
| KFD0-CS-EX1.50P | MTL5041, MTL5040 |
|