| Sensor Type |
|
| Measurement Principle |
Zirconium oxide sensor with metallic sealed reference and K Type T/C |
| Performance |
| Gas Requirements |
Typical exhaust gas |
| Measurement Range |
0.01 to 25% Oxygen |
| Accuracy |
Better than ± 2% of reading |
| Response Time (T90) |
20 seconds |
| Repeatability |
± 0.1% of reading |
| Stability |
1% per month |
| Linearity |
Better than ±1% of reading |
| Sample Flow Rate |
Flue Gas at 0.5m/s (17.65 ft/s) minimum rate |
| Maximum Sample Pressure |
Application dependent |
| Maximum Sample Temperature |
+1300°C (+2372°F) (see probe selection in
Wetted Materials) |
| Outputs |
| Output Signal |
One 0/4 to 20 mA linear, with isolated output 2nd output optional |
| Output Load |
Over 1000 Ω |
| Self-diagnostics |
Includes: oven under temperature; thermocouple separation; calibration fail; EEPROM
corruption (Diagnostic information displayed on panel meter) |
| Output Ranges |
Configurable between 0.01 to 25% |
| Alarms |
2 Alarms; User-adjustable (10 W) 1
Fault Alarm |
| Display Resolution |
0.1% in standard (0.01% on
request) |
| Power Supply |
110V (100 to 120) 50-60 Hz or 220 V (190 to 240) 50 Hz |
| Power Consumption |
110 V A |
| Ambient Temperature Range |
0 to +55°C (+32 to 131°F) |
| Sensor Temperature |
+700°C (+1292°F) |
| Operating Humidity |
5 to 90% RH non-condensing |
| Physical |
| Dimensions |
| Control unit: | 300 x 300 x 200mm (11.8 x 11.8 x 7.8 inches) |
| Sensor: | 290 x 135 x 650mm (11.4 x 5.3 x 25.6 inches) |
| Weight |
10 to 15 kg (22 to 33 lbs) depending on configuration |
| Wetted Materials |
XZR501: 304 L Stainless steel (700°C (1292°F))
XZR502: Inconel 600 (1000°C (1832°F))
XZR503: HR160 (600°C to 1000°C (1112°F to 1832°F))
XZR504: C2000 (600°C (1112°F))
XZR505: Halar coating (120°C (248°F))
XZR506: Ceramic (1300°C (2372°F))
|
| Probe Lengths |
0.4, 0.6 and 0.9m (15.75, 23.62, 35.43 in) |
| Installation |
Stack and wall-mount |
| Housing Ingress Protection |
Weather proof enclosure to sensor: IP53
Control unit: IP52 |