S8000 RS High Precision Chilled Mirror Hygrometer

The S8000 RS chilled mirror sensor directly measures the formation of condensation, giving long-term and unmatched, drift-free readings of dew-point and relative humidity. It offers a wide measurement range from -90 to +20°C dew point (-130 to +68°F). Fully automated control of the auxiliary cooling system means that no operator intervention is required even if the measured dew point changes from one end of the range to the other.


Product Features

  • Accuracy of ±0.1°C (±0.18°F)
  • Precision measurement to -90°Cdp (-130°Fdp, 100 ppb) with no need for additional cooling
  • Simple configuration and operation via touch screen interface
  • Sensor head optimized for fast response to low moisture levels
  • High measurement sensitivity
  • Microscope for visual inspection of condensate on mirror
  • Compact 19" x 4U package
  • Lightweight at 17 kg (37.5 lbs) - less than half the weight of closest competitor
  • Ethernet or USB connections
  • SD card datalogging
S8000 RS

Technical Specification

Dew-Point Sensor Performance
Measurement Technology Chilled Mirror
Measurement Range RS80: -80 to +20°Cdp (-112 to +68°Fdp)
RS90: -90 to +20°Cdp (-130 to +68°Fdp)
Measurement Accuracy* ±0.1°C (±0.18°F)
Reproducibility ±0.05°C (±0.09°F)
Mirror Gold plated copper
Temperature Measurement 4 wire Pt100, 1/10 DIN class B
Sample Flow Rate 500 to 1000 ml/min (1 to 2.1 scfh)
Sample Gas Pressure 1 MPa (10 barg/145 psig) max
Remote PRT
Temperature Measurement 4 wire Pt100, 1/10 DIN class B
Measurement Accuracy ±0.1°C (±0.18°F)
Cable Length 2m (6.6ft) (250m (820ft) max)
Flow Sensor
Measurement Range 0 to 1000ml/min (0 to 4.2 scfh)
Optional Integrated Pressure Sensor
Measurement Range 0 to 1.6 MPa (0 to 16 bara, 0 to 232 psia)
Measurement Accuracy 0.25% Full Scale
Measurement Units barg, psig, kPa, MPa
Monitor
Resolution User selectable to 0.001°C, depending on parameter
Measurement Units Moisture: °C dp or °F dp, % RH, g/m3, g/kg, ppmV, ppmW (SF6)
Temperature: °C or °F
Pressure: barg, psig, kPa, Mpa
Outputs Analog: Three channels, user selectable 4-20 mA, 0-20 mA or 0-1 V
Digital: USB and Modbus TCP (over Ethernet)
Alarm: Two volt-free changeover contacts, one process alarm, one fault alarm; 1 A @ 30 V DC
HMI 5.7" LCD with touchscreen, white on blue graphics
Data Logging SD Card (512 Mb supplied) and USB interface
Supports SD Card (FAT-32) - 32 Gb max. that allows 24 million logs or 1388 days, logging at 5 second intervals
Environmental Conditions 5 to 30°C (41 to 86°F), max 80% RH
Power Supply 85 to 264 V AC, 47/63 Hz
Power Consumption 250 VA
Mechanical Specification
Dimensions 177 x 440 x 550mm (7 x 17.32 x 21.65in) (h x w x d)
Weight 17kg (37.5lbs)
Sample Gas Circuit 316 Stainless steel
Sample Gas Connections Inlet: ¼" VCR
Outlet: ¼" Swagelok
General
Calibration 5-point in-house calibration, national standards traceable as standard
UKAS accredited calibrations optional - please consult factory


* Measurement accuracy means maximum deviation between instrument under test and corrected reference. To this must be added the uncertainties associated with the calibration system and the environmental conditions during testing or subsequent use.

Related Products

S8000 Integrale S8000 Integrale
Features an integrated sensor head allowing measurement in the range of -60°Cdp to +40°Cdp, combining the measurement sensitivity and ±0.1°C dew point accuracy of the RRS system.
QMA401 QMA401
The next generation Advanced Quartz Crystal Microbalance analyzer from Michell Instruments is designed to provide reliable, fast and accurate measurement of trace moisture content in a variety of applications where keeping moisture to a minimum is of critical importance.
S8000Remote S8000 Remote
The S8000 Remote uses a compact remote sensor design, perfect as reference instrument in environmental applications and climatic chambers. The S8000 Remote features ±0.1°C accuracy across its -40 to +90°C dew-point measurement range.
DG-1, DG-2, DG-3, DG-4 and VDS Dew-Point Generators
A range of dew-point generators offering ease of use and fast response to humidity step changes.
S4000 TRS S4000 TRS
The S4000 TRS is designed for very dry dew point measurements to -100°C (-148°F) with ±0.1°C accuracy. It features excellent stability and reproducibility across its entire measurement range, and is used in calibration and standards laboratories worldwide.
DCS DCS
The DCS system is a premium customizable rack-mounted dew-point calibration station with cooled mirror reference.
Optidew Vision Optidew Vision