S8000 Remote High Precision Chilled Mirror Hygrometer

The S8000 Remote has all the great features of the other S8000 Series chilled mirror hygrometers, but with a convenient and compact remote sensor design.
The field-proven sensor boasts upgraded temperature control for extremely accurate dew-point measurement. Coupled with the standard ±0.1°C (±0.18°F) accuracy ambient temperature sensor, it provides fundamentally derived, high precision relative humidity readings necessary for validating the results of environmental tests.



Product Features

  • Fundamental, accurate and drift-free measurement
  • Remote sensor
  • Open design allows remote sensor to be mounted into a sample flow or simply placed in an environment to be monitored
  • -40 to +90°C (-40 to +194°F) dew-point range with
    ±0.1°C (±0.18°F) accuracy
  • Data logging to USB or SD card
  • 'FAST' guarantees frost formation below 0°C (32°F
  • Sensor operates in pressures up to 20 barg (290 psig)
S8000 Remote

Technical Specification

Dew-Point Sensor Performance
Measurement Technology Chilled Mirror
Sensor TypeAcetalAluminium
Dew-Point Range-40°Cdp @ sensor temp of +20°C to +90°Cdp @ sensor temp of +90°C-20°Cdp @ sensor temp of +20°C to +90°Cdp @ sensor temp of +90°C
Operating Temperature-40 to +90°C (-40 to +194°F)
%RH Range< 0.5-100%< 4.5-100%
Accuracy±0.1°C (±0.18°F)
Response Speed 1°C/sec
Reproducibility ±0.05°C
Operating Pressure 0 to 20 barg (0 to 290 psig)
Remote PRT
Temperature Measurement 4 wire Pt100, 1/10 DIN class B
Accuracy ±0.1°C (±0.18°F)
Cable LengthForm C relay contacts (30 V, 1 A) for mirror contamination, optical and temperature measurement fault
Pre-set
Optional Remote Pressure Sensor
Measurement Range 0 to 25 bara (0 to 377 psia)
Accuracy 0.25% Full Scale
Measurement Units psia, bara, KPa or MPag
Pressure Transducer Thread 1/8” NPT
Monitor
Resolution User-selectable to 0.001 dependant on parameter
Measurement Units °C and °F for dew point and temperature %RH, g/m3, g/kg, ppmV , ppmW (SF6), for calculated humidities
Outputs Analog   3 channels, user selectable 4-20 mA, 0-20 mA or 0-1 V
Digital   PC Communications using Modbus RTU over USB
Alarm   Two volt free changeover contacts, one process alarm, one fault alarm; 1 A @ 30 V DC
HMI High definition, blue LCD
User-adjustable contrast
Menu navigation via five button keypad
Data Logging SD Card (512Mb supplied) and USB interface
SD Card (FAT-16) - 2Gb max. that allows 24million logs or 560 days, logging at 2 second intervals
Environmental Conditions -20 to +50°C (-4 to +122°F)
Power Supply 85 to 264 V AC, 47/63 Hz
Power Consumption 100 V A
Mechanical Specifications
Dimensions (Instrument) 180 x 400 x 320mm (7.1 x 15.7 x 12.5”) h x w x d
Dimensions (Sensor) ø45 x 128mm with M36 x 1.5-6g mounting thread
Weight 7.9kg (17.41lbs)
Cable Lengths 2, 5 or 10m (6.6, 16.4 or 32.8ft)
General
Storage Temperature -40 to +60°C (-40 to +140°F)
Detection SystemSingle optics detection system with auto adjustment
Calibration 4-point traceable in-house calibration as standard; UKAS accredited calibrations optional - please consult Michell Instruments

Related Products

Cooled Mirror Reference Instruments

S8000 Integrale S8000 Integrale
Features an integrated sensor head allowing measurement in the range of -60°Cdp to +40°Cdp, combining the measurement sensitivity and ±0.1°C dew point accuracy of the RRS system.
Optidew Vision Optidew Vision
The Optidew Vision is an excellent entry-level calibration reference. It comes with all the same features as the standard Optidew.
Optidew Optidew
The Optidew is Michell Instruments’ high performance industrial chilled mirror hygrometer supplied in a rugged IP66, NEMA4 wall-mount enclosure.
S4000 Remote S4000 Remote
The S4000 RC provides accurate measurement of dew-point, temperature and % RH from -85°C to +80°C dew point. The remote sensor is ideal for environmental monitoring.