| Performance | |
| Measurement technology | Quartz Crystal Microbalance |
| Range | 0.1 to 2000ppmV. Calibrated range 0.5 to 500ppmV with trend indication above and below calibration |
| Accuracy | ±5% for 0.5 to 500 ppmV or 0.1 ppmV whichever is greater |
| Repeatability | ±5% of the reading from 0.5 to 1000ppmV |
| Sensitivity | 0.1ppmV or 1% of the reading, whichever is the greater |
| Response time | <2 minutes for 63% response to a step change in either direction <5 minutes for 95% response to a step change in either direction |
| Automatic calibration | Internal moisture generator source calibrated traceable to NIST |
| Electrical output/input | |
| Analog output | 4–20mA for moisture (max load 500Ω) User-configurable range |
| Analog output resolution | 16 bit |
| Supply voltage | 110V (90 to 115) or 240V (220 to 245) switch selectable, 50-60Hz |
| Alarms | 3 Alarms: Volt free contact, rating 3A@250 VAC |
| Operating conditions | |
| Inlet pressure | 15psig (Adjusted to achieve the correct exhaust flow, up to 1500psig with optional pressure regulator) |
| Outlet pressure | Atmospheric |
| Sample flow requirement | 3.6 in³/hr or 300ml/min |
| Sample gas temperature | 0°C to 100°C / 32°F to 212°F |
| Operating environment | 15°C to 40°C / 59°F to 104°F , 90%RH max |
| Mechanical specification | |
| Display | 8.4 inch LCD touch-screen |
Dimensions (W x H x D) | 18.5 inches x 8.4 inches x 15.7 inches – allow extra 7.5
inches for clearance at back
470mm x 214mm x 400mm – allow extra 200mm for
clearance at back |
| Net weight | 33lbs / 15kg |