| Performance |
| Measurement Range (RH) |
0-100% RH |
| Measurement Range (T) |
-30 to +70°C |
Accuracy at 23°C Humidity |
<±2% RH (5–95% RH) |
Accuracy at 23°C Temperature |
±0.2°C |
| Stability – RH Sensor |
<±1% RH/year |
| Response Time - RH Sensor |
<10 sec typical (for 90% of the
step change) |
| Electrical output/input |
| Output Signal (RH) |
4–20mA, 0–10 V |
| Output Signal (T) |
3-wire 1/3 DIN Pt100 direct |
| Supply Voltage |
10–32 V DC for 4–20 mA output
15–32 V DC for 0–10 V output
24 V AC ±10% (PF211B) |
| Load Resistance |
Output 4–20 mA: Rload < (Uv-9)/0.02
Output 0–10 V: R > 1KΩ |
| Current Consumption |
20 mA max |
| Operating conditions |
Operating Temperature Probe, Housing, Storage |
-30 to +70°C |
| Mechanical specification |
| Ingress Protection |
IP54 |
| Housing Material |
ABS |
| Dimensions |
L=175mm, ø19mm |
| Weight |
184g |
| Electrical Connections |
5 or 6 wire output cable, 3m |