| Performance |
| Measurement Range (RH) |
0-100% RH |
| Measurement Range (T) |
-40 to +60°C / -40 to +140°F |
| RH Accuracy at 23°C / 73°F |
<±2% RH (10–90 % RH) |
| Temperature Accuracy |
±0.2°C ±0.36°F |
| Stability - RH Sensor |
±1% RH/year (depending on environmental conditions)
|
| Response Time - RH Sensor |
<10 sec typical (for 90% of the
step change) |
| Electrical output/input |
| Output Signal |
0–1VDC |
| Supply Voltage |
8-35 VDC |
| Operating conditions |
Operating Humidity Probe, Housing, Storage |
5-95% RH (non-condensing) |
Operating Temperature Probe, Housing Storage |
-40 to +60°C / -40 to +140°F -40 to +70°C / -40 to +160°F |
| Mechanical specification |
| Ingress Protection Sensor |
IP65 |
| Housing Material |
Molded polymer housing |
| Dimensions |
L=130, ø19mm / L=5.11", ø0.74" |
| Weight |
30g / 1.1oz without cable |
| Electrical Connections |
M12 |