| Performance |
| Measurement Range (RH) |
0-100% RH |
| Measurement Range (T) |
-20 to +80°C |
Accuracy at 23°C Humidity |
PC52: <±2% RH (10–90% RH) PC33: <±3% RH (30–80% RH) |
Accuracy at 23°C Temperature |
PC52: ±0.2°C PC33: ±0.3°C |
| Stability - RH Sensor |
±1% RH/year
|
| Response Time - RH Sensor |
<10 sec typical (90% of the
step change) |
| Electrical output/input |
| Output Signal |
4–20 mA, 0–1, 0–5, 0–10 V DC |
| Supply Voltage |
14–30 V DC (for 0-5 / 0-10 V output) 5–30 V DC (0–1 V & mA output) |
| Operating conditions |
Probe Housing Storage | -30 to +85°C -30 to +85°C -40 to +85°C |
| Ingress Protection Sensor |
IP65 (NEMA 4 level) |
| Housing Material |
Moulded polymer housing or stainless steel (ordering option) |
| Dimensions |
L=130mm, ø19mm |
| Weight |
30g without cable (moulded polymer version) |
| Electrical Connections |
M12 |