| PERFORMANCE |
| Measurement Range (dewpoint) |
-100 to +20°C dew point -148 to +68°F dew point
0–3000 ppmV
ppmV output or non-standard dew point
range must be specified at time of
order |
| Accuracy (dewpoint) |
±2°C / ±3.6°F dew point |
| Response Time |
5 mins to T95 (dry to wet) |
| Repeatability |
0.5°C / 0.9°F dew point |
| ELECTRICAL INPUT/OUTPUT |
| Output Signal |
4–20 or 0–20mA signal,
maximum load resistance 500 Ω
RS232 |
| Supply Voltage |
85 to 264 VAC, 50/60 Hz |
| Load Resistance |
Max 250 Ω @ 12 V
500 Ω @ 24 V |
| Current Consumption |
60 mA max |
| OPERATING CONDITIONS |
| Operating Temperature |
-40 to +75°C / -40 to +167°F
|
| Operating Humidity |
0–100% RH |
| MECHANICAL SPECIFICATIONS |
| Ingress Protection Sensor |
Monitor: IP65 (NEMA 12) front panel only Sensor: IP66 (NEMA 4) |
| Dimensions |
Monitor: 1/8 DIN case, 96 x 48 x 85 mm / 3.77 x 1.88 x 3.34"
including clearance (W x H x D)
Sensor: 132 x 27A/F mm / 5.13 x 1.06"
(length x diameter across hex flats)
|
| Weight |
150g / 5.3oz |
| Filter |
HDPE
|
| Electrical Connections |
Sensor cell supplied |
| Alarm |
Front panel configuration of alarm
points
|
| Sensor Cable |
0.8m / 2.6 ft supplied as standard (max 800m / 2600ft) |
| Power Cord |
2m / 6.5 ft cable supplied
|