Control Kit for the Calibration of Relative Humidity Sensors

The control kit consists of:

  • 5 phials of the same humidity value
  • 7 test strips
  • 1 control chamber for the version HKCxxC, without control chamber for version HKCxxS
  • Box dimensions: 100 x 140 x 40mm
  • Weight: 1.0kg



Accuracy

  • An accuracy of ±3% RH can be achieved if the test procedure is adhered to.



Operating conditions

  • Reference temperature: +23°C ±1°C
  • Environmental humidity stabilised ±10%
  • Set up conditions:
    • Temperature between 0 and +50°C with correction to be applied according to the instruction manual
    • Humidity between 40 and 60% RH
  • The instrumentation should be checked and the control kit must be kept in the same environment for at least 10 hours before the procedure is started
  • Shelf life of the solutions:
    • in phial: unlimited
    • in the control chamber: 2 hours