Performance | |
Measurement Technology | Thermal Conductivity sensor |
Measured Gases | Air, Ar, CH4, CO2, H2, He or N2 |
Gas Requirements | Non-condensing sample with particles <3μm |
Measurement Range | Selectable from 0-5 up to 0-100% 50-100% up to 90-100% |
Measurement Range (H2 & He) | Selectable from 0-1% up to 0-100% 50-100% up to 99-100%* |
Display Resolution | 0.1% |
Display Type | Backlit LCD |
Intrinsic Error (Accuracy) | < ±2% of span** |
Response Time (T90) | Standard < 20 seconds** |
Repeatability | ±0.2% of span |
Linearity | ±1% span |
Zero Stability | ±0.5% of span per month |
Span Stability | ±0.5% of span per month |
Sample Flow Rate | 100 to 500 ml/min (0.2 to 1.27 scfh) |
Sample Flow Rate(Ex) | 270 to 330 ml/min (0.54 to 0.66 scfh)) |
Sample Flow Effect (calibrated at 300 ml/min) | < 1% of span for flows: 100 to 500 ml/min |
Sample Pressure | Analyzers are callibrated at atmospheric pressure and must be operated with an atmospheric vent, unless otherwise stated |
Maximum Safe Pressure | 4 barg (58 psig) |
Sample Temperature | 5 to 45°C (+32 to +104°F) max |
Sample Cell Temperature | +50°C (+122°F) |
Electrical Specifications | |
Analog Inputs | 2 off 4-20 mA inputs One for an external sensor that can be displayed on the screen One to act as an active compensation for the process conditions |
Analog Outputs | 2 off 4-20 mA outputs |
Output Ranges | Primary range is set to the calibrated range of the instrument The second is user selectable |
Alarms | 2 off single pole changeover (SPCO) relays (250 V, 5 A max) |
Datalogging | The user can use the digital communications to log the output from the analyzer The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp |
Digital Communications | MODBUS RTU over RS485 Protocol |
Power Supply | 24 V DC; 1.5 A max |
Operating Conditions | |
Ambient Temperature | 0 to 40°C (32 to 104°F) |
Mechanical Specifications | |
Warm Up Time | < 25 minutes |
Stabilization Time | 5 minutes |
Dimensions | 234 x 234 x 172mm (w x d x h) |
Weight | 9.7kg |
Wetted Materials | 316 stainless steel, boroscillicate glass,
platinum, (plus O-ring) |
O-Ring Materials | Viton, Silicone or Ekraz |
Gas Connection | 1/4” NPT FEMALE GP models, 1/8” NPT FEMALE Ex & GP2 models |
Ingress Protection | IP66, NEMA 4X |
Hazardous Area Classification | |
ATEX | II 2GD Ex d IIB +H2 T6 Gb Ex tb IIIC T85°C Db IP66 |
IECEx | Ex d IIB +H2 T6 Gb Ex tb IIIC T85°C Db IP66 |
Temperature ranges for ATEX and IECEx as per O-ring type | Silicon: Ta = -40°C to +55°C Viton: Ta = -15°C to +55°C Ekraz: Ta = -10°C to +55°C |
CCSAUS | Class I, Division 1, Groups B,C,D T6 |
TC TR Ex-Cert | 1Exd IIB+H2 T6 Gb |
*Ranges 98-100% and 99-100% (for H2 and HE only) will require a flowing reference and selection of the 4-port model.
reference and selection of the 4-port model. **The typical intrinsic error and response time will be as above. Certain gas combinations and/or ranges may have different specifications. Please consult Michell Instruments for specific cases. For example, CO2 / N2 would be < 50 seconds for T90.
Note: If the 4-port version is selected then a reference gas with a flow rate of 300ml/min is required |