| Performance | |
| Measurement Technology | Quartz Crystal Microbalance |
| Range | 0.1 to 2000ppmv Calibrated range 0.5 to 700ppmV with trend indication above and below calibration range |
| Accuracy | ±10% max for 0.5 to 700 ppmV or 0.1 ppmV whichever is greater (±5% available on request) |
| Repeatability | ±5% of the reading from 0.5 to 700ppmV |
| Sensitivity | 0.1ppmV or 1% of the reading, whichever is the greater |
| Response Time | <2 minutes for 63% response to a step change in either direction <5 minutes for 95% response to a step change in either direction |
| Automatic Calibration | Internal moisture generator source calibrated traceable to NPL and NIST |
| Electrical output/input | |
| Analogue Output | 4–20 mA for moisture (max load 500Ω) User-configurable range |
Analogue Output Resolution | 16 bit |
| Supply Voltage | 110 V (90 to 115) or 240 V (220 to 245) switch selectable, 50/60Hz |
| Alarms | 3 Alarms: Volt free contact, rating 3 A @ 250 V AC |
| Operating conditions | |
| Inlet Pressure | 1 barg (Adjusted to achieve the correct exhaust flow, up to 200 barg with optional pressure regulator) |
| Outlet Pressure | Atmospheric |
| Sample Flow Requirement | 300 ml/min |
| Sample Gas Temperature | 0 to +100°C |
| Operating Environment | +15 to +40°C, 90%RH max |
| Mechanical specification | |
| Display | 8.4 inch LCD touch-screen |
| Dimensions | 470 x 214 x 400mm – allow extra 200mm for clearance at back (w x h x d) |
| Net Weight | 15kg |