Promet I.S.

 Process Moisture Analyzer
Promet Process Moisture Analyzers are heavy duty, industrial hygrometer systems for measurement of high pressure, process gases and vaporised liquids on natural gas platforms and terminals, petrochemical plants and industrial gas manufacturing facilities. Promet combines the latest in ceramic moisture sensing technologies with sample conditioning system design engineering capabilities to provide a reliable and rugged on-line instrument. The flexible modular concept of Promet allows Michell Instruments to work with the customer to select a system tailored to the customer's specific application and technical requirements.

Product Features

  • Total analyzer system tailored to specific customer requirements
  • Fully field interchangeable sensors
  • Pressure compensation of conversions to moisture content
  • Wide dew-point measurement range with display in any hygrometric unit
  • Analysis pressure up to 30 MPa
  • Sensor protected against glycol and other process borne liquid contaminants
  • ATEX certified operation in Zone 1 or 2 EExd option
Promet I.S.

Technical Specification

PROCESS MOISTURE ANALYSER
Sensor Transmet IS
Measurement Range -100 to +20°C dew point
Accuracy ±1°C from +20 to -60°C dew point
±2°C from -60 to -100°C dew point
Hygrometric Units Digital processing linearisation with pressure compensation from measured dew point with pressure compensation to any other preferred unit-dew point, Lb/mmscf and mg/Nm³ (for natural gas), ppmV
Pressure Input Fixed value (user programmed) or dynamic measurement (4-20 mA pressure transmitter, optional)
Resolution 0.1°C dp, 0.1°F dp, 0.01 ppmV, 0.1 mg/m³, 0.01 Lb/mmscf
Analogue Output 4-20mA isolated (max load 600 W)
Sensor Calibration Traceable to British (NPL) and American (NIST) National Humidity Standards
Alarms Two user adjustable for set points and operating configurations (4 optional)
Mounting Panel 19” anodised aluminium, 3U high; 132 x 200mm clearance depth
Accuracy ±0.25% FS
SAMPLE CONDITIONING SYSTEMS
Enclosure Glass fibre reinforced polyester (flame retarding, anti-static). All fixtures stainless steel. Option for complete enclosure in AISI 316 stainless steel
Dimensions 775H x 612W x 329D mm
Enclosure Mounting Stainless steel wall mounting brackets
Ingress Protection IP65 / IP66 optional (NEMA 4)
Cable/Pipework All through base of enclosure. Pre-machined M20 entries clearance holes for sensor field cable and heater power supply cable entries. 6mm bulkhead compression couplings for process gas supply and exhaust
Flow Rate 5 Nl/min (add 10 Nl/min for configuration with 'continuous drain' arrangement)
Analysis Pressure Up to 30 Mpa
Enclosure Temperature Heater/thermostat adjustable set-point range 0 to control 50°C
OPTIONS:
Optional vortex cooling option for desert conditions

Heater power supply 110/120 or 220/240/255 VAC, 50/60 Hz. Power consumption 100 or 200 W max. Add 40 W for single stage electricity heated pressure regulation where used (80 W for two stage).
Operating shaded position, on or off-shore, -40 to +40°C environment (+60°C max transient).
Interconnecting Sensor Cable

Field cable spec Polyethylene insulated instrumentation cable to BS5308 Part 1-two collectively screened dual conductors 0.5mm². Polyethylene bedding, galvanised steel wire armour, and PVC flame retardant outer sheath in blue. Low smoke and fume or halogen free outer sheath options available.
Max length 250m (greater dependent upon cable spec).
Terminating link Medium duty, grey PVC sheathed, braided cables provided for safe area termination to Promet Monitor and hazardous area termination inside the sample conditioning system enclosure.
Safety Aspects

The complete Promet package meets the requirements of BS5345: Part 1: 1989 and other European codes of practice for Zone 1 potentially hazardous area operation of the sensor and sample conditioning system to analyse gases of all gas groupings inclusive of IIC. The Promet Monitor must be located in a designated non-hazardous area (or in an optional EExd housing).
All relevant components in the potentially hazardous area are themselves individually certified to ATEX standards. Transmet IS is ATEX certified by EECS for use in hazardous areas to Ex II 1G EEx ia IIC T4 when used with a pair of isolation or shunt diode barriers for signal and power supply connections. Transmet IS has been FM listed (USA) CL1, Div1, Gp A, B, C & D, T4 and CSA (Canada) certified to Ex ia (Class 1, Groups A, B, C & D). The standard sampling system enclosure is constructed in anti-static, flame retardant GRP to EN50 014.6.3. The sample conditioning system heater and thermostat are certified EEx d IIC T5 and EEx ed IIC T6 respectively through PTB. Electrically heated pressure regulators (where required) are certified EEx d IIC T3 through ISSeP.

Applications

Natural Gas & Petrochemicals

Related Products

Condumax II Condumax II
Condumax II is a fully self-contained hydrocarbon dew point analyzer that uses the proven "Dark...
Promet EExd Promet EExd
For the last thirty years, Michell Instruments has provided on-line moisture measurement...
Cermet II I.S. Cermet II I.S.
For continuous dew point measurements in hazardous areas with safe area display, Cermet II I.S....
Cermax I.S. Cermax I.S.
Cermax I.S. provides raw performance, rugged reliability and advanced features for measuring the...
Easidew TX I.S. Easidew TX I.S.
An ATEX certified, low cost, 2-wire, rugged impedance dew-point transmitter for continuous...
Easidew PRO I.S. Easidew PRO I.S.
Intrinsically safe dew-point transmitter...