| Performance specifications  |  
							 
 
													 
							  
								| RH Measurement Range (RH) |  
								0 to 100% RH |  
							 
 
							  
								| RH accuracy @ 23°C  |  
								± 0.8% RH (5 to 95%RH)  |  
							 
 
							  
								| RH thermal coefficient  |  
								<0.03% RH/°C typical |  
							 
 
							  
								| RH measurement response time |  
								< 1 sec to RH event |  
							 
 
							  
								| RH element |  
								Michell H8000 |  
							 
 
							  
								| RH long term stability |  
								±1% RH per year |  
							 
 
							  
								| Temperature measurement range |  
								-40 to +85°C |  
							 
 
							  
								| Temperature accuracy |  
								±0.2°C |  
							 
 
							  
								| Temperature resolution |  
								±0.01°C |  
							 
 
							  
								| Temperature technology |  
								PT1000 1/3 DIN Class B |  
							 
 
							  
								| Recommended storage range |  
								+10 to +40°C |  
							 
 
							  
								| Calibration |  
								Traceable 5-point calibration certificate |  
							 
 
							  
								| Application Software |  
								HS3 Application Software downloadable from www.michell.com |  
							 
 
						
							  
								| Electrical Specifications  |  
							 
 
							  
								| Voltage output signals |  
								0-1 V, 0-2.5 V, 0-5 V, 0-10 V |  
							 
 
							 
								| Digital output signal |  
								Modbus RTU over RS485 2-wire |  
							 
							  
								| Electrical thermal coefficient |  
								0.0004%/°C |  
							 
							  
								| Load resistance |  
								0-1, 0-2.5 V : 10K Ω   0-5, 0-10 V : 50K Ω  | 
							 
							  
								| Supply voltage range |  
								5 to 28 V DC |  
							 
							  
								| Supply current consumption |  
								5V : 4mA   28V : 7mA  |  
							 
							  
								| Supply protection |  
								Protected against reverse voltage and overvoltage |  
							 
							  
								| Configurable calculated scales/ranges |  
								Dew point: -40 to +100°C, Water activity: 0 to 1, Absolute humidity: 0 to 200 g/m³,   
Specific enthalpy: 0 to 800 KJ/kg,  Frost point: -50 to +10°C	 |  
							 
							  
								| Configurable temperature scales/ranges |  
								°C, °F : -20 to 80°C, 0 to 50°C, -40 to +60°C, -30 to +70°C, 0 to 100°C |  
							 
						 
								| CE conformity |  
								2004/108/EC heavy industrial immunity |  
							 
                                            
								
							  
							  
								| Operating Specifications  |  
							 
 
							 
							  
								| Probe operating temperature |  
								Probe: -40 to +85°C   Interchangeable sensor:-40 to +120°C |  
							 
 
							 
							  
								| Recmommended Storage Temperature |  
								+10 to +40°C |  
							 
							 
							  
								| Mechanical specification  |  
							 
 
							  
								| Ingress protection |  
								IP67 (NEMA 6) |  
							 
 
                                                             
								| Material |  
								Probe & sensor body: Solid glass filled PBT Interconnect ring: Anodise aluminium  Filter assembly: HDPE or moulded polymer & PVDF |  
							
							 
								| Dimensions |  
								 Probe:L=145mm, ø15mm   Interchangeable sensor:L= 56mm, ø12mm  | 
							
                          
 
							           
								| Weight |  
								31g approx (packed weight 45g)  |  
							 
  
								
							  
								| Electrical Connector |  
								M12 5 pin (A coded)  |  
							 
							 
							  
								| Mating connector/cable (optional) |  
								M12 connector/cable 2, 5, 10 meter assembly |  
							 
							 
							  
								| Product marking |  
								Indelible laser etched alphanumeric/bar coded   identification |  
							 
							 	  
								| Maintenance kits ( Optional )   |  
							 
 
							   
								| HS3 Probe simulator kit |  
								Desk-mounted configuration kit, 3 x sensor   simulators (25%, 50%, 75% RH at 23°C), 
spare probe electronics, carry case								 |  
							 
							   
								| HS3 Probe configuration kit |  
								Desk-mounted configuration kit, spare   probe electronics, carry case  |  
							 
							   
								| HS3 Probe logger/ configuration converter cable |  
								 1.8 metre USB to RS485 converter cable with M12 connector  |