H5000 & 5100

 Capacitive Relative Humidity Sensor
The operating principle of these capacitive relative humidity sensors are based on the hygroscopic properties of their polymer, which is used like a dielectric in a capacitor. The polymer gets in equilibrium with its humid environment quickly, and reversibly, and changes its capacity value depending on the humidity level.






Product Features

  • Capacitive thin film sensor
  • Measuring range: RH 0–100%, Temp: -30 to +200°C
  • Mixing ratio: 250g water/Kg of dry air
  • Low hysteresis
  • Response time: 4 seconds
H5000 & 5100

Technical Specification

H5000
H5100
Response time
90% of scale for a step change
from 11% to 75% RH
4 sec
4 sec
Operating range
   Humidity
   Temperature
   Pressure

0–100% RH
-30 to +200°C (-22 to +392°F)
0.04–30 bar (0.6-400 psi)

0–100% RH
-30 to +100°C (-22 to +212°F)
0.04–30 bar (0.6-400 psi)
Mixing ratio
250g water / Kg dry air
Nominal capacity 75% RH @ 23°C
500 pF ± 10%
Sensitivity 11-75% RH @ 23°C
0.86 pF/% RH
Linearity 11–90% RH @ 23°C
± 2.5% RH
Long term stability (12 months) control @ 11% RH
< 1% at 23°C
Maximum air speed (without protection)
< 20m/sec
Hysteresis
Typical value = 0.5% RH
D Factor loss tangent @10 KHz 75% RH @ 23°C
Typical value = 0.007
Supply voltage
Peak-to-peak
2.5 V AC
DC component < 0.2 V
Operating frequency range
5–300 KHz
Protection cap
No
Yes
Weight
0.1g
1g