| Performance |
| Measurement Technology |
Impedance |
| Measurement Range |
-100 to +20°Cdp |
| Accuracy |
±2°Cdp |
| Run time |
12-16 hours |
| Charge time |
16 hours for maximum charge |
Electrical input/output |
| Output |
4-20 mA current maximum load
resistance 400 Ω |
| Power Supply |
4 rechargable 'C' cells with battery status
indicator |
Operating conditions |
| Operating Temperature |
-20 to +50°C |
| Storage Temperature |
-40 to +75°C |
| Operating Pressure |
0.1-2 MPa (1-20 barg/14.5-290 psig)
standard; up to 30 MPa (300 Barg/4351
psig) with optional Swagelok
® couplings |
Mechanical specification |
| Display |
Flush mounted 3.5 digit red LED |
| Case |
Yellow propylene with charger, sample tubing
and output connector stored in the lid |
| Weight |
4 kg total weight |
| Enclosure Rating |
| Case Closed |
NEMA Type 6P (submersible) |
| Case Open |
NEMA Type 4 (weatherproof) |
| Sample Connections |
Push fit pneumatic fittings for gas inlet/outlet |
| Sample Block |
Stainless steel, fully self-contained
sample system with fixed orifice ports
for flow control/pressure or atmospheric
measurement and built-in filtration using a
standard drop-in cartridge |
| Filter Cartridge |
Removes 99.5% of particles ≥ 0.3 μm
supplied with cartridge installed. Spare
cartridges are available (part no: 9996031) |
| Interchangeability |
Sensor is fully interchangeable with other
Easidew products |
| Sample Tubing |
2m of 6mm O/D PTFE supplied |