| PERFORMANCE |
| Measurement Range (RH) |
0-100% RH |
| Measurement Range (T) |
-30 to +85°C |
| Accuracy RH @ 23°C |
<±2% RH (5–95% RH) |
| Accuracy T @ 23°C |
±0.2°C (-10 to
+50°C) |
| Stability - RH Sensor |
±1% RH/year
|
| Response Time - RH Sensor |
<10 sec typical (90% of the
step change) |
| ELECTRICAL INPUT/OUTPUT |
| Output Signal (RH) |
0-1 VDC |
| Output Signal (T) |
3-wire 100 Ω 1/3 DIN direct Pt100 |
| Supply Voltage |
5 VDC ±5% |
| Load Resistance (T) |
R > 5KΩ
Digital: CMOS compatible |
| Current Consumption |
1.5mA max |
| OPERATING CONDITIONS |
| Operating Temperature |
Sensing element: -30 to +85°C Housing: -40 to +85°C Storage: -40 to +85°C |
| Operating Humidity |
Probe: 0–100% RH Housing: 0–100% RH Storage: 0–100% RH |
| MECHANICAL SPECIFICATIONS |
| Filter |
AISI 316
Stainless steel mesh |
| Housing Material |
AISI 316 |
| Electrical Connections |
Nipple M16x1 + Nut |
| Dimensions |
L=53mm, ø22mm |
| Weight |
65g |