| PERFORMANCE |
| Measurement Range (RH) |
0-100% RH |
| Measurement Range (T) |
0 to +100°C, -20 to +80°C, 0 to
+50°C |
| Accuracy RH @ 23°C |
PC52: <±2% RH (10-90% RH)
PC33: <±3% RH (30–80% RH) |
| Accuracy T @ 23°C |
PC52: ±0.2°C (-10 to
+50°C) PC33: ±0.3°C (5 to +40°C) |
| Stability - RH Sensor |
±1% RH/year
|
| Response Time - RH Sensor |
<10 sec typical (90% of the
step change) |
| ELECTRICAL INPUT/OUTPUT |
| Output Signal |
4–20 mA signal, 0–1, 0–5, 0–10 VDC PC52-RS232, RS485 |
| Supply Voltage |
14–30 VDC, 5–30 VDC (0–1 V & mA output) |
| OPERATING CONDITIONS |
| Operating Temperature |
Probe: -30 to +70°C Housing: -30 to +70°C Storage: -40 to +75°C |
| Operating Humidity |
Probe: 5-95% RH Housing: 5-95% RH Storage: 5-95% RH |
| MECHANICAL SPECIFICATIONS |
| Ingress Protection Sensor |
IP65 |
| Housing Material |
Plastic or stainless steel |
| Electrical Connections |
M12 |
| Dimensions |
L=130, ø19mm |
| Weight |
30g without cable (plastic version) |