| Performance
|
| Measurement Range: |
1-stage | -30 to +90°Cdp at sensor temperature of 20°C |
2-stage | -40 to +90°Cdp at sensor temperature of 20°C |
High temperature | -20 to 130°Cdp at sensor temperature of 20°C |
| Measurement Accuracy |
±0.2°Cdp
±0.1°C temperature
±0.15°Cdp accuracy optional |
| Measurement Units |
°C, °F dew point; % RH; °C, °F temperature; g/m³ g/kg; aw; ∆ (t-t dew point) |
| Response Speed |
1°C/sec plus settling time (dew-point dependant) |
| Power Supply |
85 to 264 V AC, 47/440 Hz |
| Dew-point Sensor
|
| Mirror options available |
Gold plated copper Solid gold 316 stainless steel - high temp sensor |
| Temperature Measurement |
4 wire Pt100, 1/3 DIN Class B |
| Sample Flow Rate |
0.1 to 2 Nl/min in sampling block |
| Max Velocity |
10 m/sec direct insertion 30 m/sec with sintered guard fitted |
| Pressure |
2 MPa (20 barg) 25 MPa (250 barg) optional |
| Ingress Protection |
IP66 (NEMA 4) IP65 25 MPa (250 barg) sensor (NEMA 12) |
| Cable Length |
2m (Maximum 30m) |
| Remote PRT |
| Temperature Measurement |
4-wire Pt100 1/10 DIN class B |
| Cable Length |
2m (Maximum 30m) |
| Transmitter Electronics |
| Resolution |
0.1 for °C, °F and % RH 0.01
for g/m³ and g/kg |
| Outputs |
| Analog: | Two channels 0/4-20 mA |
| Digital: | RS232 (RS485 optional) |
| Alarm: | Volt free contact, 2A @ 30 V DC |
| Status LEDs |
Power on, DCC and alarm status |
| Operating Temperature |
-20 to +50°C ambient |
| Enclosure |
Standing case with carry handle. Panel
mounting kit optional |
| Ingress Protection |
IP54 (NEMA 2) |
| Cable Pack |
Power and RS232 cables |