| General
|
| Measurement Range |
-60 to +40°Cdp |
| Measurement Accuracy* |
±0.1°Cdp |
| Stability |
Better than 0.1°Cdp |
| Repeatability |
<0.05°C |
| Measurement Units |
°Cdp or °Fdp; % RH; °C or °F temperature; g/m³, g/kg; ppmV; ppmW; SF6; ml/min |
| Response Speed |
2°C/sec + settling time |
| Power Supply |
85 to 264 V, 47/63 Hz; 100 VA |
| Operating Temperature Range |
-20 to +50°C |
| Dew-Point Sensor
|
| Mirror |
Gold plated copper |
| Temperature Measurement |
4 wire Pt100, 1/10 DIN class B |
| Pressure |
1.7 MPa (17 barg) (for higher pressure consult
factory) |
| Flow Sensor |
| Measurement Range |
0 to 1000 ml/min |
| Display Unit |
| Resolution |
User-selectable to 0.001 dependent on parameter |
| Outputs |
| Analog | Three channels; user-selectable 4-20 mA, 0-20
mA or 0-1 V |
| Digital | USB, SD Card slot |
| Alarm | Two volt free changeover contacts, one process
alarm, one fault alarm; 1A @ 30 V DC |
| Cable Pack |
Supply and USB cables. Output connectors suitable for indoor use
|
| Optional Remote PRT |
| Temperature Measurement |
4 wire Pt100, 1/10 DIN class B |
| Measurement Accuracy* |
±0.1°C |
| Measurement Units |
°C or °F |
| Cable Units |
2 metres (250 metres max)
|
| Optional Pressure Sensor (In-Built) |
| Measurement Range |
2.5 MPa (25 bara) |
| Measurement Accuracy* |
0.25% Full Scale, typical |
| Measurement Units |
barg; Psig; Kpg; MPa |
* |
Measurement accuracy means maximum deviation between instrument under
test and corrected reference. To this must be added the uncertainties associated
with the calibration system and the environmental conditions during testing or
subsequent use. |