| H6000 | H6100 |
Response time 90% of scale for a step change from 11% to 75% RH | 20 sec | 20 sec |
Operating range Humidity Temperature Pressure | 0–100% RH -30 to +200°C / -22 to +392°F 0.04–30 bar / 0.6-400 psi | 0–100% RH -30 to +100°C / -22 to +212°F 0.04–30 bar / 0.6-400 psi |
| Mixing ratio | 250g/8.82oz water / Kg dry air |
| Nominal capacity 75% RH @ 23°C / 73°F | 500 pF ± 10% |
| Sensitivity 11-75% RH @ 23°C / 73°F | 0.86 pF/% RH |
| Linearity 11–90% RH @ 23°C / 73°F | ± 2.5% RH |
| Long term stability (12 months) control @ 11% RH | < 1% at 23°C / 73°F |
| Maximum air speed (without protection) | < 20m/sec |
| Hysteresis | Typical value = 0.5% RH |
| D Factor loss tangent @10 KHz 75% RH @ 23°C / 73°F | Typical value = 0.007 |
Supply voltage Peak-to-peak | 2.5 VAC DC component < 0.2 V |
| Operating frequency range | 5–300 KHz |
| Protection cap | No | Yes |
| Weight | 0.1g / 0.0004oz | 1g / 0.035oz |