| Response Speed |
<1 minute for a 20 ppb moisture
event in 1 ppb base |
| Moisture Range |
1 ppb to 20,000 ppm |
| Pressure Range |
0 to 10 Barg absolute |
| Pressure Rating |
300 Barg absolute |
| Leak Tested To |
0-9 mbar l/sec-¹ |
| Operating Flow Range |
1 to 10 l/min |
| Operating Temperature |
5 to 40°C |
| Storage Temperature |
30 to 70°C |
| Power Supply |
20 to 28 VDC |
| Power Consumption |
5 W max |
| Fus Rating for any External Fuse |
500 mA Quickblow |
| Data Output |
Modbus ASCII on 2-wire RS-485 |
| Warning and Alarm Outputs |
2 open-collector transistors |
| Maximum Voltage |
40 VDC |
| Maximum Current |
100 mA |
| MATERIALS EXPOSED TO
GAS |
| Housing Material |
Stainless Steel AISI316L VAE,
electropolished <10 Ra surface finish |
| Sensor Construction |
Aluminium oxide dielectric, gold upper
plate, aluminium lower plate |
| Internal Volume |
4.4cm-³ |
| Weight |
450g |
| Gas Connections |
1/4" male VCR |
| Electrical Connections |
9-pin D-type male connector |
| Electro-Magnetic Compatibility |
EN50081-2 and EN50082-2 |