Promet EExd

 Prozessfeuchte Analysatorr
The complete moisture measurement package for critical process gas applications.



Produkt Merkmale

  • EExd Flameproof certified - ATEX and CSA
  • Complete hazardous area installation
  • Single or dual-channel measurement
  • Calibration traceable to NPL (UK) and NIST (US)
  • Remote global access options
  • Sour gas compatible
  • Field calibration maintenance
  • Range -120 to +30°Cdp, 0.001 to 30,000 ppmV
Promet EExd

Technische Spezifikation

Channel Configurations Single and two channel
Moisture Parameters Dew point °C and °F
Moisture Content Automatic pressure compensated conversions: ppmV for natural gas and ideal gas, lbs/MMscf and mg/m3 for natural gas.
Dew point at a pressure input by the user for natural gas and ideal gas.
Calculations for natural gas moisture content based on either IS0 18453 or IGT#8 to customer order preference
Analysis Pressure barg, MPa and psig
Sensor Technology Michell Ceramic Moisture Sensor
Measurement Range -120°Cdp to +30°Cdp
1 ppbV to 30,000 ppmV
0 to 250 barg
Calibration Range -100°Cdp to +20°Cdp
10 ppbV to 23,000 ppmV
Accuracy Dew point: ±1°C between -59.9 & +20°Cdp
Moisture content: ±10% of reading
Dew point: ±2°C between -60 & -100°Cdp
Moisture content: ±20% of reading
Analysis Pressure: ±0.25% FS
Measured Resolution 0.1°C: -80 to +20°Cdp
1°C: -100 to -80°Cdp
Displayed Resolution Dew point: 0.1°C
Moisture content: autoscale, 5 digits
MPa and barg: 0.1, psig: 1.0
Resolution 0.1°C between +20 and -80°Cdp
1°C between -80 and -100°Cdp
Temperature Coefficient Algorithm compensation
HMI: Keyboard Interface Capacitive touch-screen through glass
Display Vacuum fluorescent
Datalogging A rolling maximum of 150 data logs is available
Each log records time, date, moisture and pressure values for each channel
5 minutes minimum and 60 minutes maximum logging intervals can be set by the user
Communications and Output Two non-isolated 4-20 mA per measurement channel. 500 Ω maximum load. Range and parameter settable by user.
Modbus RTU @ 9600 baud-rate
Alarms: two volt free contacts per channel; one process value and one instrument status
Also available via Modbus communications
Main Unit Enclosure Internal temperature control for condensation protection and stable analysis conditions
Type Flameproof EExd
Construction Cast copper-free aluminium
Finish Chromate primer, Polyester P9010 powder coated (black). Meets BS3900
Operating Pressure Max 206 barg
Sample Connections ¼" NPT (female)
Sample Flow Rate 0.5 to 5 l/min
Sample Flow Alarm Optional
Power Supply 90 to 260 V AC, 50/60 Hz, 180 W Main Unit
Operating environment Indoor/Outdoor -20°C to +60°C
Max 95% RH Shaded location
Remote Interface
Optional for indoor safe area installation host capacity Up to 31 Michell EExd Process Instrumentation Range analyzers in any combination of models via Modbus
Display Backlit color 15cm/6" LCD
User Interface Integral mouse, on-screen keyboard
Functions View and modify analyzer operating settings
Record parameters to file and view virtual chart recorder
Analyzer diagnostics
Ports LAN, USB2
Data Download CSV (comma separated variable) format for easy import to Excel™
Remote Access OPC (OLE for Process Control) over a LAN
Imbedded web page/unique IP address for PC access
Power Supply 100-280 V, 50/60 Hz

Anwendungen

erdgas petrochemie

Verwandte Produkte

Promet I.S Promet I.S
Ein komplettes Turnkey Hygrometersystem mit eigensicherem Sensor für Feuchtemessung in...
Condumax II Condumax II
Automatische, On-line Messung des Kohlenwasserstoff-Taupunktes und des Feuchtegehaltes in Erdgas...
Cermet II I.S Cermet II I.S
Eigensicheres Online Hygrometer für kontinuierliche Taupunktmessung in explosionsgefährdeten...
MDM300 & MDM300 I.S. MDM300 & MDM300 I.S.
Ein Highspeed-Taupunkt Hygrometer für wiederholbare, schnelle und genaue Messungen der Spurenfeuchte im Feldeinsatz...
Liquidew I.S Liquidew I.S
On-line measurement of moisture content in liquids is accurate, fast and reliable with the Liquidew I.S..