| Performance |
| Measurement Range (dp) |
-100 to +20°Cdp |
| Accuracy (dp) |
±2°Cdp |
| Response Time |
5 mins to T95 (dry to wet) |
| Repeatability |
0.5°C dew point |
| Electrical output/input |
| Output Signal |
4-20 mA (2-wire connection, current source) User-configurable over range
|
| Output |
Dew point, moisture content for ppmV |
| Output Range |
Dew point: -100 to +20°C Moisture content in gas: 0 - 3000 ppmV |
| Supply Voltage |
12 - 28 V DC |
| Load Resistance |
250 Ω @ 12 V Max
500 Ω @ 24 V |
| Current Consumption |
20 mA max |
| Supply Voltage Influence |
±0.005% RH/V |
| Operating conditions |
| Operating Temperature |
-40 to +60°C
|
| Operating Humidity |
0–100% RH |
| Operating Pressure |
45 MPa (450 barg) max
|
| Flow Rate |
1 to 5 l/min mounted in standard
sampling block 0 to 10 m/sec direct
insertion |
| Temperature Coefficient |
Temperature compensated across
operating temperature range
|
| Mechanical specification |
| Ingress Protection |
IP66 in accordance with standard
BS EN 60529:1992 NEMA 4 in
protection accordance with standard
NEMA 250-2003 |
| Housing Material |
Stainless steel
|
| Dimensions |
L=132mm x ø27mm
|
| Weight |
150g |
| Filter |
HDPE Guard <10 μm
80μm sintered guard (optional)
|
| Interchangeability |
Fully interchangeable transmitters
|
Fault Conditions (factory
programmed) | Condition Sensor fault Under-range dew point Over-range dew point
|
| Output 23
mA 4 mA 20 mA |
|