| Sensor |
| Measurement Range |
-100 to +20°Cdp (-148 to +68°Fdp) |
| Accuracy (dew point) |
±1°C from -60 to +20°Cdp
(±1.8°F, -76 to +68°Fdp)
±2°C from -100 to -60°Cdp
(±3.6°F, -148 to -76°Fdp) |
| Enclosure |
Stainless steel |
| Ingress Protection |
IP66 in accordance with standard BS EN
60529:1992 and NEMA 4 in protection and
accordance with standard NEMA 250-2003 |
| Operating Temperature |
-40 to +60°C (-40 to +140°F)
compensated over -20 to +40°C (-4 to +104°F) |
| Storage Temperature |
-40 to +109°C (-40 to +165°F) |
| Operating Pressure |
From vacuum to 40 MPa
(400 barg/5801 psig) (max) |
| Flow Rate |
1 to 5 l/min (2.1 to 10.5 scfh) |
| Gas Velocity |
0 to 10m/sec (0 to 32.8 ft/sec) |
| Process Connection |
5/8" UNF parallel thread |
| Sensor Calibration |
Traceable to UKAS and NIST |
| Analyzer |
| Measurement Units |
°C, °F dew point; ppmV with user selectable range and resolution;
lb/MMSCF range dependent on pressure (natural gas); g/m³ (natural gas). Digital
processing and linearization with active or
passive pressure compensation |
| Resolution |
0.1 °C from -80 to +20°Cdp
(0.18°F, -112 to +68°Fdp)
1°C from -100 to -80°Cdp
(1.8°F, -148 to -112°Fdp) |
| Secondary Measurement Variable |
Pressure using an optional external
transducer over a range of 0-45 MPa (450
barg/6527 psig. Completely user-configurable
for any industry standard transducer with an
4-20 mA output. Automatic compensation for
ppmV, lbs/MMscf and g/m³ units. |
| Alarm Relays |
Two user adjustable isolated 10A form 'C'
rated at 240 V AC or 24 V DC |
| Outputs: |
| Analog: | 4-20 mA as standard (max load 500 Ω)
user configurable and scalable
|
| Digital: |
RS232 |
| Power Supply |
Universal 85 to 265 V AC, 50/60 Hz or 95
to 370 V DC |
| Operating conditions |
| Operating Temperature |
| Analyzer: | 0 to +50°C (32 to +122°F)
|
| Sensor: |
-20 to +50°C (-5 to +122°F) |
| Ingress Protection |
| Analyzer: | IP54 (NEMA 12) standard (front panel only)
IP66 (NEMA4) with optional cover (front panel only)
|
| Sensor: |
IP66 (NEMA4) |
| Options |
| Outputs: |
| Analog: | 0-20 mA optional (max load 500 Ω)
0-10 V optional (min load 5 K Ω)
both user configurable and scalable
|
| Digital: |
RS485: provides 2-way communication - full
data access to user and total instrument
configuration |
| Alarm Relays |
1 or 2 additional 5 A relays (max 4 total) n/o
type fully user configurable and adjustable over
the full range, full choice of units and sensor
fault alarm functions |
| Power Supply |
18 to 36 V AC; 9 to 60 V DC |
| Cable |
Standard 2m (6.56ft) (1200m/3937ft max) |